Uses:
The device uses physical method to achieve disinfection and
sterilization. It takes advantage of high-temperature and hot air to
cause intoxication on microbial oxidation, protein denaturation, and
electrolyte concentration. By disinfection principle, it destroys
protoplasmic cell to kill all the microorganisms in a certain period
of heating time.
Characteristics:
1. Cold rolling steel electrostatic spraying exterior.
2. Stainless steel inner chamber; with the semicircular arcs at
corners for easy cleaning. The space between the shelves in the
chamber is adjustable. The two sides heating air passage and bottom
heating cover can be fast assembly and disassembly.
3. PID Microprocessor digital controller has the function of timing,
lack of water and over-temperature protection.
4. The special air forced heat radiation construction ensure the low
noise fan temperature less than 50°C. Turbine centrifugal fan ensure
the fan can be used more longer time.
5. High quality heat isolation door with wide hollow tempered glass
observation window, adjustable air-tight buckle lock and silicon
rubber seal.
BE function type has the following
characteristics other than the above functions:
1. With adjustor to adjust the fan
speed, the user can choose use the fan and adjust the speed ensure
the high temperature evenness, or not use the fan to avoid the
powder or other small samples move in the inner chamber.
2. Large LCD screen shows the multiple data, has the function of
timing and over-temperature protection.
3. Independent audible and visible temperature-timing alarm system,
cut off the power when over-temperature limited automatic ensures
experiments run safety and no accident would happen.
4. Can be connected the RS485 connector and software to download and
save all the data via RS485 into computer.
(Optional Accessories)
Optional accessories:
a. Desk printer
b. Change the far infrared heating tube becomes to the far infrared
Hot-air Sterilizer
Note: model SRX230 has no function of speed
adjustment.
B series are standard
type with LED screen; BE series are function type with LCD screen.
Main parameters
Model |
SRX30B/BE |
SRX45B/BE |
SRX65B/BE |
SRX85B/BE |
SRX125B/BE |
SRX230B/BE |
Voltage |
220V/50Hz |
Tem. Range (°C) |
RT+5~300°C |
Tem. Motion (°C) |
+/-1°C |
Tem. Resolution Ratio (°C) |
0.1°C |
Ambient tem. (°C) |
5-40°C |
Heating power (w) |
0.8 |
1.2 |
1.6 |
1.8 |
2.3 |
3 |
Inner Chamber Size (mm) |
310*310*310 |
350*350*350 |
400*360*450 |
450*420*450 |
500*450*550 |
600*500*750 |
Exterior Size (mm) |
400*510*695 |
500*550*735 |
550*550*840 |
590*630*830 |
636*680*915 |
730*670*1220 |
Packing Size (mm) |
610*540*830 |
660*590*875 |
670*640*970 |
680*720*920 |
760*750*1070 |
900*800*1350 |
Load per rack (Kg) |
15/kg 2 for standard configuration |
N.W. (Kg) |
39 |
42 |
48 |
52 |
58 |
70 |
G.W. (Kg) |
52 |
55 |
61 |
65 |
73 |
115 |
Timing range |
1-9999min |
|