Air Velocity Meter

EMF, RF & Radiation Meter

solar power meter

CO Meter

CO2 Meter

Temperature & Humidity Meter

lux meter

Sound Level Meter, Analyzer & Calibrator

Insulation Tester

UV Light Meter

Standard accessories:

UM-4(AAA)1.5V battery 6 PCS

users manual

USB cable (TM-802)

AC100~240V DC9V/0.5A(9mm)1 (TM-802)

Install CD disk (TM-802)

TM-801 & TM-802 CO Meter

Features:
- Backlit 4 digit LCD with maximum 1000 counts.
- Sample rate: 1 time/second
- Data logging capacity: 8000 records, USB interface. (TM-802)
- Max/Min/Average Hold.
- Data Hold.
- Low battery indication
- Auto power off with disable function.
- Adjustable alarm.
- Sample rate: 1 time/second.
Specifications:

(At 23C0.5C, RH<75%)

CO Measurement:

Range

0 ~ 1000ppm

Resolution

1ppm

Accuracy

5% + 2ppm

Repeatability

< 2%

Response time

< 30 seconds.

Zero drift in long term

< 5ppm

Sensitivity drift

< 5% (per year)

Typical sensor life

3 years

Temperature Measurement:

Range

-20~50C/ -4~122F

Resolution

0.1C/ 0.1F

Accuracy

1.0C/ 1.8CF

Power supply: 1.5V UM-4(AAA) battery 6 PCS or AC100~240V or DC Adaptor 9V/0.5A(9mm).
Battery life: approx 90 hours.
Operating Temperature & Humidity: -20C to +50C,
Storage Temperature & Humidity : -10C to +50C, <70% RH.
Dimension: 1567335mm(LxWxH).
Weight: approx 250g (Batteries not included).

 