PROVA-11
AC/DC mA True RMS Clamp Meter
|
|
|
Range |
Output |
Accuracy |
DCA 0-4000mA |
1mA |
+/- 1.5%rdg +/- 5dgts |
¡¡ |
DCA 0-30A |
10mA |
+/- 1.5%rdg +/- 3dgts |
¡¡ |
ACA 0-400mA |
0.1mA |
+/- 1.5%rdg +/- 5dgts |
True RMS |
ACA 0-4000mA |
1mA |
+/- 1.5%rdg +/- 8dgts |
True RMS |
ACA 0-30A |
10mA |
+/- 2.0%rdg +/- 5dgts |
True RMS |
DCV 0-400V |
0.1V |
+/- 1.0%rdg +/- 3dgts |
¡¡ |
ACV 0-400V |
0.1V |
+/- 2.0%rdg +/- 5dgts |
True RMS |
Power Source: 2 X 1.5V AA UM3
Batteries
Current Consumption:
10mA
Operating Condition:
-10 ºC to 50 ºC at 85% max. RH
|
|
|
|
|
|
|
Range |
Resolution |
Accuracy |
Remarks |
DCA 0-40A |
10mA |
+/- 1.0% +/- 2dgts |
¡¡ |
DCA 0-150A |
100mA |
+/- 1.0% +/-2 dgts |
¡¡ |
DCA 150-200A |
100mA |
-2.2% +/- 2dgts |
¡¡ |
ACA 0-40A |
10mA |
+/- 1.0% +/- 3 dgts |
¡¡ |
ACA 0-150A |
100mA |
+/- 1.0% +/- 3 dgts |
¡¡ |
ACA 150-200A |
100mA |
-2.2% +/- 3dgts |
¡¡ |
DCV 0-400V |
0.1 V |
+/- 1.0% +/- 2dgts |
¡¡ |
ACV 0-400V |
0.1 V |
+/- 1.0% +/- 2 dgts |
¡¡ |
W 0-400 W |
0.1 W |
+/- 1.0% +/-2dgts |
beep if < 40 W
(Approx.) |
Hz 0-1000KHz |
0.1-100 |
+/- 0.5% +/- 2dgts |
1V sensitivity Auto
Range |
|
Power
Source: 2 X 1.5V AA UM3 Batteries
Current Consumption: 10mA
Operating Condition: -10 ºC
to 50 ºC at 85% max. RH |
|
|
|
PROVA-5601
Clamp-on Ground Resistance Tester
|
|
|
-
Non-Contact Ground
Resistance Measurement
-
No Auxiliary
Electrode Needed
-
Jaw Diameter ¦µ 23 mm
(5601)
-
0.025 ¨C 1500 Ohm
|
-
0.2mA¨C
30A(5601),True RMS
-
116 Sets of Memory
-
Programmable Hi and
Lo Alarm
-
Programmable Data
Logging
|
|
-
Telecommunication
-
Cable TV system
-
Electrical Power
Plant
|
-
Cellular Phone
System
-
Factory and Facility
-
Gas Station Tank
|
|
¡¡ |
Ground
Resistance (Auto range) |
Range |
Resolution |
Accuracy of Reading
1, 2 |
0.025 - 0.250 ¦¸ |
0.002 ¦¸ |
¡À 1.5% ¡À 0.05 ¦¸ |
0.250 - 1.000 ¦¸ |
0.02 ¦¸ |
¡À 1.5% ¡À 0.05 ¦¸ |
1.001 - 9.999 ¦¸ |
0.02 ¦¸ |
¡À 1.5% ¡À 0.1 ¦¸ |
10.00 - 50.00 ¦¸ |
0.04 ¦¸ |
¡À 2.0% ¡À 0.3 ¦¸ |
50.01 - 99.99 ¦¸ |
0.04 ¦¸ |
¡À 2.0% ¡À 0.5 ¦¸ |
100.0 - 200.0 ¦¸ |
0.4 ¦¸ |
¡À 3.0% ¡À 1.0 ¦¸ |
200.1 - 400.0 ¦¸ |
2 ¦¸ |
¡À 5.0% ¡À 5 ¦¸ |
400.1 - 600.0 ¦¸ |
5 ¦¸ |
¡À 10% ¡À 10 ¦¸ |
600.1 - 1500 ¦¸ |
20 ¦¸ |
¡À 20% |
1 Loop
resistance noninductive, external field <
200 A/m, external electrical field < 1 V/m,
conductor centered.
2 Resistance
Measurement Frequency:
1.667KHz, 3.333KHz(optional) |
|
Ground/Leakage Current (Auto range, 50/60
Hz, True RMS, Crest Factor < 3.5)
|
Range |
Resolution |
Accuracy of Reading |
0.200 - 1.000 mA |
0.001 mA |
¡À 2.0% ¡À 0.05 mA |
1.00 - 10.00 mA |
0.01 mA |
¡À 2.0% ¡À 0.03 mA |
10.0 - 100.0 mA |
0.1 mA |
¡À 2.0% ¡À 0.3 mA |
100 - 1000 mA |
1 mA |
¡À 2.0% ¡À 3 mA |
0.20 - 4.00 A |
0.001A |
¡À 2.0% ¡À 0.03A |
4.00 - 30.00 A(5601) |
0.01A |
¡À 3.0% ¡À 0.03A |
|
Accuracy of
Resistance Calibration Plate |
¡À0.5% |
Data Logging Capacity |
116 records |
Data Logging Interval |
1 to 255 seconds |
Conductor Size |
23mm /0.90"(5600,5601) |
Jaw Opening |
27mm /1.06"(5600,5601) |
Battery Type |
9V IEC 6 LR61
(Alkaline) |
Display Type |
4 digits 9999 counts
LCD |
Overload Indication |
OL |
Battery Life |
3000 measurements |
Range Selection |
Auto |
Power Consumption |
40mA |
Sampling Time |
0.5 seconds |
Operating Temperature |
0ºC to 50ºC (14ºF to
122ºF ) Operating Humidity: Less than
85% RH |
Storage Temperature |
-20ºC to 60ºC (-4ºF
to 122ºF ) Storage Humidity: Less than
75% RH |
Dimension( 5600, 5601
) |
257mm (L) x 100mm(W)
x 47mm(H),10.1" (L) x 3.9"(W)
x 1.9"(H) |
Weight( 5600, 5601) |
640g/1.41lbs |
Accessories |
Resistance
Calibration Plate x 1 |
Battery 9V
(Installed) x 1 |
Carrying Box x 1 |
Users Manual x 1 |
|
CM-02/04
Clamp
Meter/Automotive Clamp
|
|
-
DC/AC 0.01A
Resolution
-
AC/DC
0.01A~200A
-
23mm Jaw
Diameter
-
One Touch
Zero
-
Temperature
and Capacitance
-
Auto Range
V. W .Hz. o C.
-
3 3/4 Digits
(4000 Counts)
-
0~9999
RPM(CM-04)
|
|
|
|
|
AC Voltage (Autorange, Input
Impedance 5M W )
|
|
DC Voltage (Auto range, Overload
Protection 800VAC )
|
|
Frequency (Hz, Autorange &
Sensitivity selected by range
button) |
|
Resistance( W )&Audible
Continuity (Open Circuit Voltage
0.5V): |
|
|
Resistance( W )(Autorange &
Manual, Open Voltage<=0.5V)
|
|
Capacitance (Autorange & Manual)
|
|
TEMP K TYPE ºC (Autorange &
Manual, ºF optional):
|
|
Conductor Size |
23mm
Diameter |
Battery |
1.5V
AA UM3 Battery x 2 |
Display |
3
3/4Diguts (4000Counts)
|
Range Selection |
Auto
and Manual |
Power Consumption |
10mA
(Approx) |
Sampling Time |
2Time/sec (Digital) 20Time/sec
(Bar graph) |
Operating Temperature |
0 ºC
~ 50 ºC (32 ºF ~122 ºF)
|
Operating Humidity |
LESS
than 80% RH |
Dimension |
7.2"(L) x 2.5"(W) x 1.4"(H)
183mm(L) x 63.6mm(W) x 35.6mm(H)
|
Weight |
190g. (Batteries Included)
|
Accessories |
Adaptor (for K-type Thermocouple
and Capacitor) x 1
User Manual x 1
5V Battery x 2
K-type Thermocouple x 1
Carrying Bag x 1 |
Option |
RPM
Adaptor(COM04) x 1 |
|
|
|
|
CM-03
Leakage Current Tester
|
|
|
-
10 m A Ultra High Resolution
-
Large Conductor Diameter (30mm)
-
ACA 5 Ranges (40mA, 400mA, 4A, 40A, 60A)
-
It can measure up to 400A, but over 60A, the
measuring result has not been calibrated.
|
|
|
Range
|
Resolution
|
Accuracy
|
50/60Hz |
Wide
|
ACA 0-40mA
|
10 uA
|
+/- 1.0% rdg +/- 3 dgts
|
+/- 2.0% rdg +/- 5 dgts (40 to 1KHz)
|
ACA 0 -400mA
|
100 uA
|
+/- 1.0% rdg +/- 3 dgts
|
+/- 2.0% rdg +/- 5 dgts (40 to 1KHz)
|
ACA 0-4A
|
1mA
|
+/- 1.0% rdg +/- 3 dgts
|
+/- 2.0% rdg +/- 5 dgts (40 to 1KHz)
|
ACA 0-40A
|
10mA
|
+/- 1.0% rdg +/- 3 dgts
|
+/- 2.0% rdg +/- 5 dgts (40 to 1KHz)
|
ACA 0-60A(0-50A)
|
0.1A
|
+/- 1.5% rdg +/- 3 dgts
|
+/- 3.0% rdg +/- 5 dgts (40 to 1KHz)
|
ACA 0-60A(50-60A)
|
0.1A
|
+/- 3.0% rdg +/- 5 dgts
|
+/- 3.5% rdg +/- 5 dgts (40 to 1KHz)
|
ACV 0-400V
|
0.1V
|
+/- 1.5% rdg +/- 2 dgts
|
+/- 2.0% rdg +/- 4 dgts (40 to 1KHz)
|
W 0-400 Ohm
|
0.1 Ohm
|
+/- 1.0% rdg +/- 2 dgts
|
Beef if < 40 W
|
|
Power Source:
2 X 1.5V SUM-3 Batteries
¡¡ |
Operation
Temperature/Humidity:
-10 ºC to 50 ºC at 85% max.
|
|
|
|
CM-05 DC/AC Current Probe
|
|
|
-
Accurate and High Resolution
-
23 mm Conductor Diameter
-
One Touch Zero
|
|
|
Range |
Output |
Sensitivity |
Accuracy |
DCA 0-4A
|
DC 100mV/A
|
10mA/mVDC
|
+/- 2.0% +/- 0.005A
|
DCA 0-40A
|
DC 10mV/A
|
0.1A/mVDC
|
+/- 1.0% +/- 0.02A
|
DCA 0-150A
|
DC 1mV/A
|
1A/mVDC
|
+/- 1.0% +/- 0.2A
|
DCA 150-200A
|
DC 1mV/A
|
1A/mVDC
|
-2.2% +/- 0.2A
|
ACA 0-4A
|
AC 100mV/A
|
10mA/mVAC
|
+/- 2.0% +/- 0.003A
|
ACA 0-40A
|
AC 10mV/A
|
0.1A/mVAC
|
+/- 1.5% +/- 0.03A
|
ACA 0-150A
|
AC 1mV/A
|
1A/mVAC
|
+/- 1.5% +/- 0.3A
|
ACA 150-200A
|
AC 1mV/A
|
1A/mVAC
|
-2.2% +/- 0.3A
|
|
Power Source:
2 X 1.5V
AA UM3 Batteries
Current Consumption:
8mA
Operating Condition:
-10 ºC to 50 ºC at 85% max.
RH
|
|
|
|
CM-07 TRMS DC/AC Clamp Meter
|
|
|
-
High Resolution DC 10mA, AC 1mA
-
True RMS
-
Non-contact Current Frequency Measurement
|
|
|
Range |
Resolution |
Accuracy |
Remarks |
DCA 0-40A
|
10mA
|
+/- 1.0% +/- 2dgts
|
¡¡
|
DCA 0-150A
|
100mA
|
+/- 1.0% +/- 2dgts
|
¡¡
|
DCA 150-200A
|
100mA
|
- 2.2% +/- 2dgts
|
¡¡
|
DCA 200-400A
|
100mA
|
- 4.0% +/- 4dgts
|
¡¡
|
ACA 0-4A
|
1mA
|
+/- 1.5% +/- 3dgts
|
True RMS
|
ACA 0-40A
|
10mA
|
+/- 1.0% +/- 3dgts
|
True RMS
|
ACA 0-150A
|
100mA
|
+/- 1.0% +/- 3dgts
|
True RMS
|
ACA 150-200A
|
100mA
|
- 2.2% +/- 3dgts
|
True RMS
|
ACA 200-400A
|
100mA
|
- 4.0% +/- 4dgts
|
True RMS
|
Hz 0-1000KHz
|
0.1-100Hz
|
+/- 0.5% +/- 2dgts
|
0.1A sensitivity Auto Range
|
|
Power Source:
2 X 1.5V AA UM3
Batteries
Current Consumption: 10mA
Operating Condition: -10 ºC
to 50 ºC at 85% max. RH
|
|
|
PROVA-15 AC/DC mA Current Probe
|
|
|
-
High Resolution DC 1mA, AC 0.1mA
-
Connect to an Oscilloscopes to observe Current
Waveform
-
Connect to Chart Recorder/Datalogger/ DMM to
Record Current Data
|
|
|
Range |
Output |
Sensitivity |
Accuracy |
DCA 0-400mA
|
DC 1mV/A
|
1mA/mV DC
|
+/- 2.0% +/- 1mV
|
DCA 0-4A
|
DC 100mV/A
|
10mA/mV DC
|
+/- 1.0% +/- 0.5mV
|
DCA 0-30A
|
DC 10mV/A
|
0.1A/mV DC
|
+/- 1.0% +/- 0.3mV
|
ACA 0-400mA
|
AC 1mV/A
|
1mA/mV AC
|
+/- 2.0% +/- 0.5mV
|
ACA 0-4A
|
AC 100mV/A
|
10mA/mV AC
|
+/- 1.5% +/- 0.8mV
|
ACA 0-30A
|
AC 10mV/A
|
0.1A/mV AC
|
+/- 1.5% +/- 0.3mV
|
|
Power Source:
2 X 1.5V AA UM3
Batteries
Current Consumption: 10mA
Operating Condition: -10 ºC
to 50 ºC at 85% max. RH
|
|
|